HOME |
PRODUCTS |
TECH SUPPORT |
ABOUT US |
CONTACT US |
View Cart
Checkout
SEARCH
Advanced Search
PRODUCT
CATEGORIES Load Cells / Force Sensors Beam Load Cells Torque Sensors Load Cell Displays
Signal Conditioning Data Acquisition
Software Load Cell Accessories
TECH
SUPPORT
Online Manuals
Wiring / Color Code
Calibration Certificates
Load Cell Troubleshooting
Warranty / Repair
Load Cell Terminology
Calculators
Free Load Cell Catalog
Track Your Order
Learn about Plug and Play Smart Sensors
We gladly accept:
View Cart
Checkout
12 BIT DATA ACQUISITION BOARD
WITH 16 INPUT AND 2 ANALOG OUTPUTS
MODEL DAQ-16122-PCMCIA
(NI DAQCard 6024E)
Features
Two 12-bit analog outputs, 8 digital
I/O lines, two 24-bit counters
NI-DAQ driver software to simplify
configuration and measurements
Four analog input signal ranges
NIST-traceable calibration certificate
Digital triggering
SPECIFICATIONS
Bus
Analog
Inputs
Resolution
Sampling
Rate S/s
Input
Range
Analog
Outputs
Resolution
Output
Rate
Output
Range
Digital
I/O
Counter/
Timers
Triggers
PCMCIA
16 SE/8/DI
12 bits
200 kS/s
±.05
to ±10V
2
12 bits
10 kS/s
±10V
8
2, 24-bit
Digital
10 kS/s
maximum when using the single ended DMA channel for for analog
ouput. 1 kS/s maximum when using the single DMA channel for
either for either analog input or counter / timer operation.
ANALOG
INPUT ACCURACY SPECIFICATIONS
Nominal
Range (V)
Absolute
Accuracy
Relative
Accuracy
%
of
Reading
Offset
(mV)
Noise
+ Quantizatiom (mV)
Temp
Drift
(%/C)
Absolute
Accuracy
at Full Scale (mV)
Resolutin
(mV)
Positive
FS
Negative
FS
24
Hrs
1
year
Single
Pt.
Averaged
Single
Pt.
Averaged
10.0
-10.0
.0872
.0914
8.830
3.910
1.042
.0010
19.012
5.890
1.370
5.0
-5.0
.0272
.0314
4.420
1.950
.521
.0005
6.517
2.950
.686
.05
-.05
.0872
.0914
.462
.195
.052
.0010
.972
.516
.069
.05
-.05
.0872
.0914
.066
.063
.007
.0010
.119
.073
.009
Note: Accuracies are
valid for measurements following internal calibration. Averaged
numbers assume dithering and averaging of 100 single readings.
Measurements accuracies are listed for operational temperatures
within ±1 degree C of internal calibration temperature
and ± 10 degrees of external or factory calibration temperature.
One year calibration interval recommended. The Absolute Accuracy
at Full Scale calculations were performed for a maximum range
input voltage, 10 (for example, 10V range for ± 10V range)
after one year, assuming 100 pt averaging of data.
Order
Code Model
Capacity
Range
Description
Price
Quantity
Add to
Cart
Home |
Products |
Tech Support |
About Us |
Contact Us