12 BIT DATA ACQUISITION BOARD
WITH 16 INPUT AND 2 ANALOG OUTPUTS
MODEL DAQ-16122-USB
(NI DAQPad-6020E)
Features
Two 12-bit analog outputs, 8 digital
I/O lines, two 24-bit counters
NI-DAQ driver software to simplify
configuration and measurements
Four analog input signal ranges
NIST-traceable calibration certificate
Digital triggering
Four analog input signal ranges
SPECIFICATIONS
Bus
Analog
Inputs
Resolution
Sampling
Rate S/s
Input
Range
Analog
Outputs
Resolution
Output
Rate
Output
Range
Digital
I/O
Counter/
Timers
Triggers
USB
16 SE/8/DI
12 bits
100 kS/s
±.05
to ±10V
2
12 bits
20 S/s
±10V
8
2, 24-bit
Digital
10 kS/s
maximum when using the single ended DMA channel for for analog
ouput. 1 kS/s maximum when using the single DMA channel for
either for either analog input or counter / timer operation.
ANALOG
INPUT ACCURACY SPECIFICATIONS
Nominal
Range (V)
Absolute
Accuracy
Relative
Accuracy
%
of
Reading
Offset
(mV)
Noise
+ Quantizatiom (mV)
Temp
Drift
(%/C)
Absolute
Accuracy
at Full Scale (mV)
Resolutin
(mV)
Positive
FS
Negative
FS
24
Hrs
1
year
Single
Pt.
Averaged
Single
Pt.
Averaged
10.0
-10.0
.072
.076
6.380
3.467
.846
.0010
14.826
5.729
1.114
5.0
-5.0
.019
.021
3.198
1.733
.423
.0005
4.6710
2.865
.557
2.5
-2.5
.072
.076
1.608
.867
.211
.0010
3.7190
.1.432
.278
1.0
-1.0
.072
.076
.653
.347
.085
.0010
1.4980
.573
.111
Note: Accuracies are
valid for measurements following internal calibration. Avearged
numbers assume dithering and averaging of 100 single readings.
Measurements accuracies are listed for operational temperatures
within +-1 degree C of internal calibration temperature and
± 10 degrees of external or factory calibration temperature.
One year calibration interval recommended. The Absolute Accuracy
at Full Scale calculations were performed for a maximum range
input voltage, 10 (for example, 10V range for ± 10V range)
after one year, assuming 100 pt averaging of data.