16 BIT DATA ACQUISITION BOARD WITH 16 INPUT AND 2 ANALOG OUTPUTS |
| |
MODEL DAQ-16162-FW |
| |
| (NI DAQPad-6052E) |
 |
| Features |
- Two 16-bit analog outputs, 8 digital
I/O lines, two 24-bit counters
- NI-DAQ driver software to simplify
configuration and measurements
- Four analog input signal ranges
- NIST-traceable calibration certificate
- Digital triggering
| |
|
|
|
|
| SPECIFICATIONS |
| FW |
16 SE/8/DI |
16 bits |
333 kS/s |
±.05
to ±10V |
2 |
16 bits |
333 kS/s |
±10V |
8 |
2, 24-bit |
Analog &
Digital |
| 10 kS/s
maximum when using the single ended DMA channel for for analog
ouput. 1 kS/s maximum when using the single DMA channel for
either for either analog input or counter / timer operation. |
| ANALOG
INPUT ACCURACY SPECIFICATIONS |
| 10.0 |
-10.0 |
.035 |
.037 |
947 |
981 |
87 |
.00006 |
4.747 |
1145 |
115 |
| 5.0 |
-5.0 |
.005 |
.007 |
476 |
491 |
43.5 |
.00001 |
.876 |
573 |
57.3 |
| 2.5 |
-2.5 |
.035 |
.037 |
245 |
245 |
21.7 |
.00006 |
1.190 |
286 |
28.6 |
| 1.0 |
-1.0 |
.035 |
.037 |
99 |
98 |
8.7 |
.00006 |
.479 |
115 |
11.5 |
| Note: Accuracies are
valid measurements following internal calibration. Averaged
numbers assume dithering and averaging of 100 single readings.
Measurements accuracies are listed for operational temperatures
within ± 1 degree C of internal calibration temperature
and ± 10 degreesc of external or factory calibration
temperature. One year calibration interval recommended. The
Absolute Accuracy at Full Scale calculations were performed
for a maximum range input voltage, 10 (for example, 10V range
for ± 10V range) after one year, assuming 100 pt averaging
of data. |
|
|
|