|
| |
|
UNAMPLIFIED TRANSDUCERS |
| SYMPTOM/PROBLEM: |
|
ACTION |
| No Output: |
|
Verify correct wiring. |
| |
|
Verify excitation per calibration sheet. |
| |
|
Make sure pressure, load, etc. is being applied. |
| |
|
Check strain gage bridge for continuity per
Transducer
Checkout Guide (please scroll down). |
| Erratic/Intermittent: |
|
Check electrical connection for discontinuity or damage. |
| Output or Zero Drift: |
|
Check for isolation resistance between bridge wiring
and transducer body. |
| |
|
Make sure pressure, load, etc. is constant. |
| |
|
Check stability of excitation power supply. |
| |
|
Check millivolt output with voltmeter. |
| |
|
Check for RFI/EMI interference. |
| |
|
Make sure there are no rapid changes in temperature. |
| Incorrect Output: |
|
Check actual input and output resistance per Transducer Checkout Guide (please scroll down) for possible change or open bridge. |
| |
|
Check zero offset to see if high. |
| High Zero Offset: |
|
Usually indicates transducer was over ranged beyond specifications. Over ranged transducers should be
recalibrated. |
| |
|
Check actual input and output resistance measurements
per Transducer
Checkout Guide (please scroll down) for
possible
changes or open bridge. |
| |
|
|
| |
|
TRANSDUCER WITH INSTRUMENT READOUT |
| SYMPTOM/PROBLEM: |
|
ACTION |
| Erratic Display: |
|
Check electrical connection and wiring code for pin layout. |
| No Display/No Output Voltage: |
|
Check power line fuse per instrument instructions. |
| Blinking Display: |
|
Indicated overload; make sure wires are all connected and
transducer is within its range. |
| Incorrect Readout: |
|
Check transducer range on label. |
| |
|
Verify that system was set-up per instructions. Review
set-up procedure. |
| |
|
Refer to Transducer Checkout Guide (please scroll down) and verify that transducer operates properly. |
| |
|
Use Resistor Shunt-Calibration to verify calibration when integrated with Instrument Readout, or connect resistor
across (-) Excitation and (-) Signal. Refer to Certificate of Calibration for correct resistor value. |
| |
|
|
|
|
| PRODUCT SERIES: CLC, CLP, DSM, GS0, HSW, LBC, LBM, LB0, LP0, LPU, LW0, MLC, MLP,
RTS, SLB, SSM, SB0, STS, SW0, THA, THB, THC, THD, TLL,TRS, TRT |
| |
| TRANSDUCER BRIDGE RESISTANCE |
| INSTRUMENT REQUIRED: Digital volt meter (DVM) set to Ohm
scale. |
| Record resistance values: |
|
|
| TEST CONDITION |
BETWEEN |
MEASURED READING |
TOLERANCE OHMS |
| 1 |
Red & Black |
_______________ |
348-353 |
| 2 |
White & Green |
_______________ |
348-353 |
| 3 |
Red & White |
_______________ |
260-266 |
| 4 |
White & Black |
_______________ |
260-266 |
| 5 |
Black & Green |
_______________ |
260-266 |
| 6 |
Green & Red |
_______________ |
260-266 |
| |
|
|
| |
If all measured resistance readings fall within listed
tolerance, proceed to the SHORT TO GROUND step (please see below) |
|
| |
If measured resistance readings at TEST COND. 1 and 2 are in the range of 700 Ohms, this would indicate an open bridge. A reading in the range of 1000 Ohms at any of TEST COND. 3 through 6 would also indicate an open leg of the bridge. |
|
| |
the result is an open bridge, see instructions on obtaining
service under warranty or non-warranty (please see below). |
|
| |
|
|
| |
SHORT TO GROUND |
|
| |
INSTRUMENT REQUIRED: Digital volt meter (DVM) set to Mohm scale. |
|
| |
Short all transducer wires to one probe from (DVM),
and touch the other probe to the transducer body. |
|
| |
NOTE: If the transducer is anodized aluminum, try touching the
probe to the threads
for a bare metal contact |
|
| |
If the measured resistance is infinite proceed
to ZERO OFFSET step
(please see below). |
|
| |
If the measured resistance is between 10k Ohms and
200K Ohms, the transducer
bridge could have gathered moisture. |
|
| |
TIP: Try placing the transducer in a convection oven at 100 degrees
F for one hour.
Remove and re-measure resistance to ground. |
|
| |
If the measured resistance reads between 1 Ohm and
5 Ohms, continuity within the bridge
is evident. See instructions on obtaining service under warranty or non-warranty
(please see below).
|
|
| |
|
|
| |
ZERO OFFSET |
|
| |
INSTRUMENT REQUIRED: Digital volt meter (DVM) set to millivolt scale 10
VDC power supply |
|
| |
Connect the Red and Black transducer
leads to the 10 VDC (5 VDC on SLB-25 and SLB-50) power supply. |
|
| |
Red (+ Excitation) to 10 VDC (5
VDC on SLB-25 and SLB-50) |
|
| |
Black (- Excitation) to return ground |
|
| |
Connect the Green and White transducer leads to the digital
voltmeter (DVM) |
|
| |
If the DVM reads higher than +.300 millivolts or lower
than -.300 millivolts, the transducer
has been over-ranged beyond specifications. |
|
| |
TIP: Most transducer instrumentation can balance out a zero offset
of + 2.000
millivolts. However, an over-ranged transducer should be re-calibrated.
See instructions on obtaining service under warranty or non warranty
(please see below). |
|
| |
|
|
|
|
|
Any of our products which, under normal operating conditions, proves defective
in material or in workmanship within one (1) year from the date of shipment
by Transducer Techniques, will be repaired or replaced free of charge
provided that you obtain a return material authorization from Transducer
Techniques and send the defective product, transportation charges prepaid
with notice of the defect, and establish that the product has been properly
installed, maintained, and operated within the limits of rated and normal
usage. Replacement product will be shipped F.O.B. our plant. The terms
of this warranty do not extend to any product or part thereof which, under
normal usage, has an inherently shorter useful life than one year. The
replacement warranty detailed here is the Buyer's exclusive remedy, and
will satisfy all obligations of Transducer Techniques, whether based on
contract, negligence, or otherwise. Transducer Techniques is not responsible
for any incidental or consequential loss of damage which might result
from a failure of any Transducer Techniques, product. This express warranty
is made in lieu of any and all other warranties, expressed or implied,
including implied warranty of merchantability or fitness for particular
purpose. Any unauthorized disassembly or attempt to repair voids this
warranty.
Advance authorization is required prior to the return to Transducer Techniques.
Before returning the item(s) either write to the Repair Department
c/o
Transducer Techniques, 42480 Rio Nedo, Temecula, CA 92590, or call (951)
719-3965 with: 1) a part number; 2) a serial number for the defective
product; 3) a technical description of the defect; 4) a no-charge purchase
order number (so products can be returned to you correctly); and,
5) ship
to and bill to addresses. Shipment to Transducer Techniques shall be
at Buyer's expense and repaired, or replacement items will be shipped
F.O.B.
our plant in Temecula, CA. Non-verified problems or defects may be subject
to a $75 evaluation charge. Please return the original calibration
data
with unit.
Advance authorization is required prior to the return to Transducer Techniques.
Before returning the items, either write to the Repair Department
c/o
Transducer Techniques, 42480 Rio Nedo, Temecula, CA 92590, or call (951)
719-3965 with: 1) a model number; 2) a serial number for the defective
product; 3) a technical description of the malfunction; 4) a purchase
order number to cover Transducer Techniques, repair cost; and, 5)
ship
to and bill to addresses. After the product is evaluated by Transducer
Techniques, we will contact you to provide the estimated repair costs
before proceeding. The minimum evaluation charge is $75. Shipment to
Transducer Techniques shall be at Buyer's expense and repaired items
will be shipped
to you F.O.B. our plant in Temecula, CA. Please return the original calibration
data with unit.
All repairs of Transducer Techniques, products are warranted for a period
of 90 days from the date of shipment. This warranty applies only to those
items which were found defective and repaired; it does not apply to products
in which no defect was found and returned as is, or merely recalibration.
Out of warranty product may not be capable of being returned to the exact
original specifications or dimensions.
|
|
| |
|
UNAMPLIFIED TRANSDUCERS |
| SYMPTOM/PROBLEM: |
|
ACTION |
| No Output: |
|
Verify correct wiring. |
| |
|
Verify excitation per calibration sheet. |
| |
|
Make sure pressure, load, etc. is being applied. |
| |
|
Check strain gage bridge for continuity per
Transducer
Checkout Guide (please scroll down). |
| Erratic/Intermittent: |
|
Check electrical connection for discontinuity or damage. |
| Output or Zero Drift: |
|
Check for isolation resistance between bridge wiring
and transducer body. |
| |
|
Make sure pressure, load, etc. is constant. |
| |
|
Check stability of excitation power supply. |
| |
|
Check millivolt output with voltmeter. |
| |
|
Check for RFI/EMI interference. |
| |
|
Make sure there are no rapid changes in temperature. |
| Incorrect Output: |
|
Check actual input and output resistance per Transducer Checkout Guide (please scroll down) for possible change or open bridge. |
| |
|
Check zero offset to see if high. |
| High Zero Offset: |
|
Usually indicates transducer was over ranged beyond specifications. Over ranged transducers should be
recalibrated. |
| |
|
Check actual input and output resistance measurements
per Transducer
Checkout Guide (please scroll down) for
possible
changes or open bridge. |
| |
|
|
| |
|
TRANSDUCER WITH INSTRUMENT READOUT |
| SYMPTOM/PROBLEM: |
|
ACTION |
| Erratic Display: |
|
Check electrical connection and wiring code for pin layout. |
| No Display/No Output Voltage: |
|
Check power line fuse per instrument instructions. |
| Blinking Display: |
|
Indicated overload; make sure wires are all connected and
transducer is within its range. |
| Incorrect Readout: |
|
Check transducer range on label. |
| |
|
Verify that system was set-up per instructions. Review
set-up procedure. |
| |
|
Refer to Transducer Checkout Guide (please scroll down) and verify that transducer operates properly. |
| |
|
Use Resistor Shunt-Calibration to verify calibration when integrated with Instrument Readout, or connect resistor
across (-) Excitation and (-) Signal. Refer to Certificate of Calibration for correct resistor value. |
| |
|
|
|
|
| PRODUCT SERIES: CLC, CLP, DSM, GS0, HSW, LBC, LBM, LB0, LP0, LPU, LW0, MLC, MLP,
RTS, SLB, SSM, SB0, STS, SW0, THA, THB, THC, THD, TLL,TRS, TRT |
| |
| TRANSDUCER BRIDGE RESISTANCE |
| INSTRUMENT REQUIRED: Digital volt meter (DVM) set to Ohm
scale. |
| Record resistance values: |
|
|
| TEST CONDITION |
BETWEEN |
MEASURED READING |
TOLERANCE OHMS |
| 1 |
Red & Black |
_______________ |
348-353 |
| 2 |
White & Green |
_______________ |
348-353 |
| 3 |
Red & White |
_______________ |
260-266 |
| 4 |
White & Black |
_______________ |
260-266 |
| 5 |
Black & Green |
_______________ |
260-266 |
| 6 |
Green & Red |
_______________ |
260-266 |
| |
|
|
| |
If all measured resistance readings fall within listed
tolerance, proceed to the SHORT TO GROUND step (please see below) |
|
| |
If measured resistance readings at TEST COND. 1 and 2 are in the range of 700 Ohms, this would indicate an open bridge. A reading in the range of 1000 Ohms at any of TEST COND. 3 through 6 would also indicate an open leg of the bridge. |
|
| |
the result is an open bridge, see instructions on obtaining
service under warranty or non-warranty (please see below). |
|
| |
|
|
| |
SHORT TO GROUND |
|
| |
INSTRUMENT REQUIRED: Digital volt meter (DVM) set to Mohm scale. |
|
| |
Short all transducer wires to one probe from (DVM),
and touch the other probe to the transducer body. |
|
| |
NOTE: If the transducer is anodized aluminum, try touching the
probe to the threads
for a bare metal contact |
|
| |
If the measured resistance is infinite proceed
to ZERO OFFSET step
(please see below). |
|
| |
If the measured resistance is between 10k Ohms and
200K Ohms, the transducer
bridge could have gathered moisture. |
|
| |
TIP: Try placing the transducer in a convection oven at 100 degrees
F for one hour.
Remove and re-measure resistance to ground. |
|
| |
If the measured resistance reads between 1 Ohm and
5 Ohms, continuity within the bridge
is evident. See instructions on obtaining service under warranty or non-warranty
(please see below).
|
|
| |
|
|
| |
ZERO OFFSET |
|
| |
INSTRUMENT REQUIRED: Digital volt meter (DVM) set to millivolt scale 10
VDC power supply |
|
| |
Connect the Red and Black transducer
leads to the 10 VDC (5 VDC on SLB-25 and SLB-50) power supply. |
|
| |
Red (+ Excitation) to 10 VDC (5
VDC on SLB-25 and SLB-50) |
|
| |
Black (- Excitation) to return ground |
|
| |
Connect the Green and White transducer leads to the digital
voltmeter (DVM) |
|
| |
If the DVM reads higher than +.300 millivolts or lower
than -.300 millivolts, the transducer
has been over-ranged beyond specifications. |
|
| |
TIP: Most transducer instrumentation can balance out a zero offset
of + 2.000
millivolts. However, an over-ranged transducer should be re-calibrated.
See instructions on obtaining service under warranty or non warranty
(please see below). |
|
| |
|
|
|
|
|
Any of our products which, under normal operating conditions, proves defective
in material or in workmanship within one (1) year from the date of shipment
by Transducer Techniques, will be repaired or replaced free of charge
provided that you obtain a return material authorization from Transducer
Techniques and send the defective product, transportation charges prepaid
with notice of the defect, and establish that the product has been properly
installed, maintained, and operated within the limits of rated and normal
usage. Replacement product will be shipped F.O.B. our plant. The terms
of this warranty do not extend to any product or part thereof which, under
normal usage, has an inherently shorter useful life than one year. The
replacement warranty detailed here is the Buyer's exclusive remedy, and
will satisfy all obligations of Transducer Techniques, whether based on
contract, negligence, or otherwise. Transducer Techniques is not responsible
for any incidental or consequential loss of damage which might result
from a failure of any Transducer Techniques, product. This express warranty
is made in lieu of any and all other warranties, expressed or implied,
including implied warranty of merchantability or fitness for particular
purpose. Any unauthorized disassembly or attempt to repair voids this
warranty.
Advance authorization is required prior to the return to Transducer Techniques.
Before returning the item(s) either write to the Repair Department
c/o
Transducer Techniques, 42480 Rio Nedo, Temecula, CA 92590, or call (951)
719-3965 with: 1) a part number; 2) a serial number for the defective
product; 3) a technical description of the defect; 4) a no-charge purchase
order number (so products can be returned to you correctly); and,
5) ship
to and bill to addresses. Shipment to Transducer Techniques shall be
at Buyer's expense and repaired, or replacement items will be shipped
F.O.B.
our plant in Temecula, CA. Non-verified problems or defects may be subject
to a $75 evaluation charge. Please return the original calibration
data
with unit.
Advance authorization is required prior to the return to Transducer Techniques.
Before returning the items, either write to the Repair Department
c/o
Transducer Techniques, 42480 Rio Nedo, Temecula, CA 92590, or call (951)
719-3965 with: 1) a model number; 2) a serial number for the defective
product; 3) a technical description of the malfunction; 4) a purchase
order number to cover Transducer Techniques, repair cost; and, 5)
ship
to and bill to addresses. After the product is evaluated by Transducer
Techniques, we will contact you to provide the estimated repair costs
before proceeding. The minimum evaluation charge is $75. Shipment to
Transducer Techniques shall be at Buyer's expense and repaired items
will be shipped
to you F.O.B. our plant in Temecula, CA. Please return the original calibration
data with unit.
All repairs of Transducer Techniques, products are warranted for a period
of 90 days from the date of shipment. This warranty applies only to those
items which were found defective and repaired; it does not apply to products
in which no defect was found and returned as is, or merely recalibration.
Out of warranty product may not be capable of being returned to the exact
original specifications or dimensions.
|
|